8
ORT 4 ORT 5
Wiring 3 cable glands with integral cable clamps PG 11, 3 cable glands with integral cable clamps PG 11,
screw-type terminals with wire guard, screw-type terminals with wire guard,
max. conductor size 1.5 mm² max. conductor size 1.5 mm²
Mains supply 230 V ± 10 %, 50/60 Hz 230 V ± 10 %, 50/60 Hz
115 V ± 10 %, 50/60 Hz (option) 115 V ± 10 %, 50/60 Hz (option)
124 V ± 10 %, 50/60 Hz (option) 124 V ± 10 %, 50/60 Hz (option)
Rated voltage see name plate Rated voltage see name plate
Fuse Thermal fuse M 0.1 A Thermal fuse M 0.2 A
Power consumption 15 VA 25 VA
Measured quantity Turbidity Turbidity
Measuring range 0 – 25 ppm Range I: 0 – 50 TU/F*)
Measuring range see name plate Range II: 0 – 25 ppm*)
Measuring range see name plate
Lamp voltage for light emitter No-load voltage 14 V No-load voltage 14 V
with light emitter connected: < 12 V AC, 10 W with light emitter connected: < 12 V AC, 10 W
Actual-value output 0 – 1 mA 0 – 20 mA (option 4 – 20 mA)
load 0 ... 120 Ω load 0 ... 500 Ω
Output 1 volt-free relay contact 2 volt-free relay contact
Max. contact rating with a switching Max. contact rating with a switching
voltage of 24/115/230 V AC: voltage of 24/115/230 V AC:
resistive 4 A, inductive 0.75 A at cos ϕ = 0.5 resistive 4 A, inductive 0.75 A at cos ϕ = 0.5
Max. contact rating with a switching Max. contact rating with a switching
voltage of 24 V DC: 4 A voltage of 24 V DC: 4 A
Life or relay: 10 x 10
6
switching cycles Life of relay: 10 x 10
6
switching cycles
Contact material: AgCdO 5 µ hard-gold plated Contact material: AgCdO 5 µ hard-gold plated
Switchpoint Adjustable over complete measuring range Adjustable over complete measuring range
Delay of response 0 s with external wire linke, 10 s 0.5 – 15 min. adjustable, see name plate
without wire link, see name plate
Indicators and adjustors 1 LED
ALARM, 1 LED OPERATION, 1 LED ALARM (INSTANTANEOUS), 1 LED ALARM (DELAYED),
1 LED
MALFUNCTION, 1 potentiometer for switchpoint, 1 LED OPERATION, 1 LED MALFUNCTION,
2 potentiometers for adjustment 0/100 % 1 LED analogue display
RESOLUTION 3 %,
1 potentiometer for switchpoint,
2 potentiometers for adjustment 0/100 %,
1 potentiometer for adjustment of delay of response,
1 change-over switch for measuring range I/II,
1 change-over switch for switchpoint/measured value
Protection IP 54 to DIN 40050 IP 65 to DIN 40050
Max. admissible ambient 0 – 55°C 0 – 55°C
temperature
Housing material Sheet steel, hammer enamel finish Die-cast aluminium, blue enamel finish
Weight 1.6 kg 5.4 kg
Dimensions See Fig. 1 See Fig. 1
*) TU/F = turbidity units of Formazin suspension, ppm = parts per million
If the condensate contains suspended
solids (scratching of the glass) installation
as per Fig. 5 should be chosen. If this does
not help and/or a large amount of flash
steam escapes install a separator
upstream of the sensor (Fig. 8)
1. Lid and base can be screwed
turned through 90°.
2. Screw in nipples , screwed unions
and ball valves supplied with the
sensor (use teflon tape or hemp for
sealing).
3. They have to be fitted into the lid and
base or directly at the inlet, outlet or
purging point.
4. The screwed unions facilitate
positioning of the ball valves so that the
levers can be arranged in one plane
(front).
L
M
Screw in vent nipple when the fluid
temperature is above the max. ambient
temperature (Fig. 2).
When the fluid temperature is below the
ambient temperature remove the plastic lid
from the drying cartridge (blue colour)
and screw in the drying cartridge. (Fig. 2)
If the fluid temperature is very high provide
a non-insulated pipe upstream of ORG 12/
ORG 22 which will allow the fluid to cool
down to 120 °C when reaching the sensor.
The second inlet at the lid must be open
for the rinsing water or the calibrating fluid.
In the presence of large amounts of gas
the tapping point must be at the bottom of
the line – use a welding pocket in
accordance with DIN 2618 (Fig. 7, Pos.*)
In the case of very contaminated
condensate: downward flow of the fluid
through the glass cylinder
ORG 12/ORG 22
Install and wire ORG 12/ORG 22 in
accordance with the examples given in
figures 4 to 9.
Mount ORG 12/ORG 22 in a bypass at a
lower level than the main line so that any air
and flash steam can pass through the main
line.
To ensure a flow through the bypass a
throttling point should be provided in the
main line, e. g. GESTRA non-return valve
type RK 66.
Tap product line at the side from the
bottom to prevent air and dirt from
flowing into the bypass.
Any pressure drop upstream of the sensor
which might lead to the release of gases or
to flashing must be avoided.
Installation
F K
C
E D
F
K
Technical Data ORT