OR 52-5
OR 52-6
Installation Instructions 810731-01
Oil & Turbidity Detector OR 52-5, OR 52-6
Seitenansicht 0
1 2 3 4 5 6 ... 47 48

Inhaltsverzeichnis

Seite 1

OR 52-5OR 52-6Installation Instructions 810731-01Oil & Turbidity Detector OR 52-5, OR 52-6

Seite 2 - Component Parts

10Explanatary NotescontinuedTechnical Data1)At a flowrate of 2 l/min and V-shaped flow through the sensor with a pipe length of 1 m (DN 10) and 4 bend

Seite 3

11Explanatary NotescontinuedTechnical DatacontiunedName plate GL

Seite 4 - Contents

12Explanatary NotescontinuedTechnical DatacontinuedMeasuring Transducer ORT 6Inputs Directly transmitted light (D), scattered light (S)Outputs1 voltag

Seite 5

13InstallationMeasuring Sensor ORG 12, ORG 22It is highly advisable to have the installation work done by qualified staff in the manner described by t

Seite 6 - Important Notes

14InstallationcontinuedMeasuring Sensor ORG 12, ORG 22continuedSensor installed in a sample line, e. g. water treatment, beer filtering, permeate down

Seite 7

15InstallationcontinuedMeasuring Sensor ORG 12, ORG 22continuedInstallation of sensor in a condensate line with large amount of flash steamSchematic l

Seite 8

16InstallationcontinuedMeasuring Sensor ORG 12, ORG 22continuedThe cover flange and lower flange of the measuring sensor can be screwed turned thr

Seite 9

17InstallationcontinuedMeasuring Transducer ORT 6Unscrew cover screws and swing up cover in order to gain access to the holes for mounting the equipme

Seite 10 - Explanatary Notes

18WiringMeasuring Sensor ORG 12, ORG 22Before installing the measuring sensor cut off power supply!Connect light emitter and receiver via connecto

Seite 11

19WiringcontinuedToolsScrewdriver for cross-recess screws, size 1 and 2Screwdriver for slotted screws, size 2.5, completely insulated to VDE 0680Co

Seite 12

2Component PartsABDIJKLMNOQRSTUVWHPXYIGFEZCP

Seite 13 - Installation

20WiringcontinuedWiring Diagram ORG 12, ORG 22, ORT 6 with three-way valve0/4–20mAR≤500ΩDisconnecting switchMeasuring transducer ORT 6Measuring sensor

Seite 14

21Functional ElementsMeasuring TransducerThe push buttons have the following functions:= Program= Increase= Decrease= Enter/Test modeALARM12ppmseccal0

Seite 15

22Commissioning ProcedureFactory set default valuesThe measuring transducer features the following factory settings:Limit value: Alarm 1: 3 ppmLimit

Seite 16

23Commissioning ProcedurecontinuedFirst Commissioning1. Switch on measuring transducer (disconnecting switch).2. Rinse measuring transducer ORG 12/22

Seite 17

24Commissioning Procedurecontinued100% CalibrationFit turbidity standard 20 ppm supplied with the equipment to the light emitter . Unscrew union nu

Seite 18

25Commissioning ProcedurecontinuedOil / Turbidity CurvesA light beam shines through the liquid and any foreign matter which is not dissolved scat-ters

Seite 19

26Commissioning ProcedurecontinuedAdjusting Limit Alarm 11. Press , the yellow P-LED lights up.2. Use to change the LED display until LED Alarm 1

Seite 20

27Commissioning ProcedurecontinuedAdjusting time delay for alarm 11. Press , the yellow P-LED lights up.2. Use to change the LED display until LED

Seite 21 - Functional Elements

28Commissioning ProcedurecontinuedIndicating decimal pointThe 7-segment display can show a decimal point.This setting does not have any effect on the

Seite 22 - Commissioning Procedure

29OperationStartApply power.The 7-segment display shows briefly the version number of the software.Then the indicator counts until the actual value is

Seite 23

3KeyNipple with threaded endsBall valve ¾“ BSP (DIN ISO 228)Upper cover flangeLight receiver with tubeBall valve ¾“ BSP (DIN ISO 228) for purgingVent

Seite 24

30MaintenanceFunctional Test1. Push cleaning device slowly into mid-position to simulate a malfunction.2. The LED "Malfunction" lights up

Seite 25

31MaintenancecontinuedIndication of Error CodeIn the event of a malfunction the yellow LED Malfunction lights up, the relay output Malfunction opens a

Seite 26

32MaintenancecontinuedError Code TablecontinuedError codePossible reason RemedyE03Glow lamp in light emitter defective. A frequent occurrenceof this p

Seite 27

33MaintenancecontinuedError Code TablecontinuedReplacing Glass Cylinder 1. Close ball valves for inlet and outlet. Open ball valve for rinsing .2.

Seite 28

34MaintenancecontinuedCleaning Glass Cylinder1. Move ring of the cleaning device up and down. Note that dependent on the adjusted time delay Alarm

Seite 29 - Operation

35MaintenancecontinuedReplacing Glow Lamp1. Unscrew and remove union nut (outside) of tube of light emitter .2. Replace glow lamp (12 V 10 W) with

Seite 30 - Maintenance

36MaintenancecontinuedSpare PartsA spare parts kit containing the following items is available on request:ItemDesignation Qty.Glass cylinder 1O-ring 2

Seite 31

37AnnexDeclaration of Conformity

Seite 32

38AnnexcontinuedDeclaration of Conformitycontinued

Seite 33

39AnnexcontinuedDeclaration of Conformitycontinued

Seite 34

4ContentsPageImportant NotesUsage for the intended purpose...6Safe

Seite 35

40AnnexcontinuedCertificate of Type Approval

Seite 36

41AnnexcontinuedCertificate of Type Approvalcontinued

Seite 37 - Declaration of Conformity

42AnnexcontinuedCertificate of Type Approvalcontinued

Seite 38

43AnnexcontinuedCertificate of Type Approvalcontinued

Seite 39

44AnnexcontinuedType Approval Certificate

Seite 40 - Certificate of Type Approval

45AnnexcontinuedUpdatesIndexDescription Date00 New installation manual 04-08-2003

Seite 43

GESTRA Gesellschaften • GESTRA Companies • Sociétés GESTRA • Sociedades Gestra • Società GESTRAVertretungen weltweit • Agencies all over the world

Seite 44 - Type Approval Certificate

5ContentsPageOperationStart ...

Seite 45 - Description Date

6Important NotesUsage for the intended purposeUse oil & turbidity detector OR 52 only for monitoring transparent liquids to detect any ingress of

Seite 46

7Important NotescontinuedWarningDangerIf the measuring transducer and sensor are used in an inexpert or improper manner by unqualified staff, they can

Seite 47

8Explanatary NotesScope of supply1 Measuring transducer ORT6,1 Measuring sensor ORG12 or ORG 22 (depending on order),3Ball valves,3 Screwed unions

Seite 48

9Explanatary NotescontinuedFunctionThe measuring sensor is a photometric measuring device consisting of a light source (light emitter ) and a light r